Setting Measurement Limits; Figure 2-23 Limits Checking Applications - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

2
General Power Meter Functions

Setting Measurement Limits

You can configure the power meter to detect when a measurement has
crossed over a predefined upper and/or lower limit value.
Limits are boundaries set for a certain power range and it can be applied
to power, ratio or difference measurement.

Figure 2-23 Limits checking applications

In this application a swept frequency signal is applied to the input of the
Device Under Test. The power meter measures the output power. The
limits have been set at +4 dBm and +10 dBm. A fail occurs each time the
output power is outside these limits as shown in
Figure 2-24 Limits checking results
62
Swept Source
OUT
Amplitude
+10 dBm
o
o
o
+4 dBm
Device
Under Test
OUT
Fail
o
o
o
o
Fail
Frequency
N1913/1914A EPM Series Power Meters User's Guide
Power Meter
Figure 2-
24.
Upper limit line
Lower limit line