Performance Verification on the ENA
Agilent N4373A Lightwave Component Analyzer, First Edition
1 Turn on the network analyzer and the optical test set.
2 Allow the Lightwave Component Analyzer to warm up,
that is until the LEDs for "LD Temp Ready" and "PD
Temp Ready" are lighting.
3 Turn on the laser by turning the key to the upright
position.
The "On" LED lights to indicate the laser is on.
4 If your Lightwave Component Analyzer has more than one
wavelength, select the wavelength at which you want to
measure by pressing the button under the Wavelength
LEDs.
5 Perform an electrical calibration, as described in
"Calibrating the network analyzer before measurements"
on page 43.
6 Load the LCA macros.
a Press the "Macro Setup" button on the front panel of
the network analyzer.
b Select [Load & Run].
c Select LCASYSTEM.
7 Allow the Lightwave Component Analyzer to warm up,
that is until the LEDs for "LD Temp Ready" and "PD
Temp Ready" are lighting.
Getting Started
23