Performance Verification On The Ena - Agilent Technologies N4373A User Manual

Lightwave component analyzer
Table of Contents

Advertisement

Performance Verification on the ENA

Agilent N4373A Lightwave Component Analyzer, First Edition
1 Turn on the network analyzer and the optical test set.
2 Allow the Lightwave Component Analyzer to warm up,
that is until the LEDs for "LD Temp Ready" and "PD
Temp Ready" are lighting.
3 Turn on the laser by turning the key to the upright
position.
The "On" LED lights to indicate the laser is on.
4 If your Lightwave Component Analyzer has more than one
wavelength, select the wavelength at which you want to
measure by pressing the button under the Wavelength
LEDs.
5 Perform an electrical calibration, as described in
"Calibrating the network analyzer before measurements"
on page 43.
6 Load the LCA macros.
a Press the "Macro Setup" button on the front panel of
the network analyzer.
b Select [Load & Run].
c Select LCASYSTEM.
7 Allow the Lightwave Component Analyzer to warm up,
that is until the LEDs for "LD Temp Ready" and "PD
Temp Ready" are lighting.
Getting Started
23

Advertisement

Table of Contents
loading

Table of Contents