Figure 5-3 Reference Calibration Factor Pop-Up Window - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

Figure 5-3 Reference Calibration Factor pop-up window

5 Press
Now, zero and calibrate the power meter/sensor combination as follows:
6 Press
7 Connect the power sensor to the POWER REF output.
8 Press
You can reduce the steps required to carry out the zero and calibration procedure as
N O T E
follows:
1 Connect the power sensor to the POWER REF output.
2 Press
Now, set the sensor calibration factor for the frequency of the signal you
want to measure.
9 Check the current reference calibration factor setting by pressing
Does this setting match the value for the sensor? (The power sensor reference calibration
N O T E
factor can normally be found above the calibration factors table on the power sensor body.)
N1913/1914A EPM Series Power Meters User's Guide
%
to complete the entry.
and the channel
displayed.
and the channel
routine. The Calibrating pop- up is then displayed.
Zero + Cal
and
Zero + Cal
Zero + Cal A
,
Offset
,
. The value is displayed on the Cal Fac field.
Using 8480 Series Power Sensors
Zero
softkey. The Zeroing pop- up is
Cal
softkey to start the calibration
. (For dual channel meters, press
Zero + Cal B
or
as required).
5
133