Saving/Recalling Ripple Limit Table; Example Of Ripple Limit Configuration - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Even if the E5071C's sweep range is changed after the stimulus
value has been set, the stimulus value is not susceptible.

Example of ripple limit configuration

The individual frequency bands for the ripple test can overlap
each other; in this case, the ripple limit test is performed for
each frequency band.
Even if the E5071C's span value is set to zero, you must enter a
parameter for both Begin Stimulus and End Stimulus.
If the data format is Smith chart or polar, the limit test is
performed for the main response value among the two marker
response values.

Saving/Recalling Ripple Limit Table

The ripple limit table can be saved in a file and recalled later for use on the
screen. The file is saved in the csv format (with the extension *.csv), and
values are saved as a character string with the unit. The csv formatted file
can also be reused in spreadsheet software made for PCs.
Follow the steps below to save/recall the ripple limit table. This operation
should be done by using the external keyboard and/or mouse.
1. Display the ripple limit table.
2. Click Export to CSV File from the Edit Ripple Limit menu to open the
dialog box. At this time, CSV Files (with the extension *.csv) is
selected as the file type.
Measurement
323

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents