Agilent Technologies E5071C Manual page 259

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
Line/Match
1. Click Line/Match.
2. Click
softkey
Softkey
x-y
Line/Match
x-y Fwd
(Syx)
x-y Rvs
(Sxy)
3.
4. Click Return.
5. Click Done to finish TRL 2-port calibration. At this point, the
calibration coefficient is calculated and saved. The error correction
function is automatically turned on.
258
to execute Line/Match calibration.
Function
Executes line/match calibration for test ports x and y.
Use this to perform forward-direction measurement
and reverse-direction measurement at the same time.
This is equivalent to measuring both x-y Fwd (Syx) and
x-y Rvs (Sxy).
Executes line/match calibration for test ports x and y.
Use this to perform forward-direction measurement
only.
Executes line/match calibration for test ports x and y.
Use this to perform reverse-direction measurement
only.