Agilent Technologies E5071C Manual page 648

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

N4691A
N4691B
N4692A
N4696A
N4696B
Firmware A.09.10 or above, enable users to calibrate down to 9
kHz using N4431A (4-port).
The 9 kHz to 300 kHz performance is valid only for the E5071C
with Firmware revision A.09.10 or above using N4431B.
When Ecal is used with options 2D5, 4D5, 2K5 and 4K5 and in
the following condition, the waiting time is added automatically.
Firmware A.09.10: The maximum frequency of your ECal is
over 13 GHz.
Firmware A.09.20 and above: The maximum frequency of
your ECal is over 13 GHz and the stop frequency on your ENA
set over 8.5 GHz. (The waiting time is NOT added when stop
frequency is set below 8.5GHz.)
For scalar mixer calibration, the stop frequency of calibration
port is used to determine the waiting time instead of the sweep
stop frequency.
The total waiting time of each calibration type (approximately) is as
follows:
Calibration Type
1-port
2-port
3-port
4-port
ENA does not support the following ECal:
Model Name
N4693A
3.5 mm
3.5 mm
2.92 mm
7 mm
7 mm
Total waiting time for E5071C (sec)
9.1
18.1
27.1
36.1
Connector Type
2.4 mm
Product Information
10 MHz to 26.5 GHz
300 kHz to 26.5 GHz
10 MHz to 40 GHz
10 MHz to 18 GHz
300 kHz to 18 GHz
Frequency Range
10 MHz to 50 GHz