Agilent Technologies E5071C Manual page 546

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

-102
Syntax error
-310
System error
T
No.
Message
40
Target value not
found
242
TDR option
requires
advanced
system
calibration
-124
Too many digits
-223
Too much data
control is enabled.
A command or data type that is not
recognized exists.
One of the errors designated as
"system errors" in this instrument has
occurred.
Description
This error occurs when the target is
not found during the marker search
analysis after specifying the target
and executing the
CALC:MARK:FUNC:EXEC and
CALC:FUNC:EXEC commands. This
error occurs also when the bandwidth
is not found after executing the
bandwidth marker command,
CALC:MARK:BWID:DATA?
TDR requires the system information
PCC2.00 and above. This instrument
should be sent to service centre for
adjustment (with the firmware Rev
9.50 and above) to suit E5071C-TDR
option.
The number of digits of the argument
of the decimal numeric-value-data
element exceeds 255 with the
preceding 0 removed (see 7.7.2.4.1,
IEEE488.2).
The block-, expression-, or character-
string-type program data that has
been received conforms with the
standard. But it exceeds the amount
that can be processed under the
condition of the memory or
conditions specific to memory-related
devices. In this instrument, this error
occurs when the number of
Product Information