168
data not
allowed
C
No.
Message
240
Calibration
data lost
22
Calibration
method
not
selected
-
Character
148
data not
allowed
-
Command
100
error
60
Continuous
switching
may
damage
source
attenuator
D
position where this instrument does not accept
one.
Description
This error occurs when a file containing the
system calibration data is not found or in a
damaged state at time of the startup of this
instrument, indicating a failure of this instrument.
Contact an Agilent Technology sales office or the
company from which you bought the instrument.
This error occurs when the command for
validating the calibration data,
SENS:CORR:COLL:SAVE or
SENS:CORR:OFFS:COLL:SAVE is executed before
the command for selecting a calibration type,
SENS:CORR:COLL:METH:xxxx or
SENS:CORR:OFFS:COLL:METH:xxxx, is executed.
This error is not generated by front key
operations.
A character data element (not violating the
standard) has been received at a position where
this instrument does not accept one. Double
quotes (") are omitted where it is necessary to
place a parameter in double quotes ("), for
example.
A comprehensive syntax error has occurred
showing that this instrument cannot detect a
more detailed error. This code simply shows that
a command error defined in 11.5.1.1.4,
IEEE488.2 has occurred.
This error occurs when different power ranges are
selected in multiple channel measurement
settings to avoid source attenuator damage.
This error occurs in case of options
2D5, 4D5, 2K5 and 4K5.