and reverse-direction measurement at the same time.
Executes line/match calibration for test ports 2 and
2-4 Fwd
4.Use this to perform forward-direction measurement
(S42)
only.
Executes line/match calibration for test ports 2 and
2-4 Rvs
4.Use this to perform reverse-direction measurement
(S24)
only.
Executes line/match calibration for test ports 3 and 4.
3-4
Use this to perform forward-direction measurement
Line/Match
and reverse-direction measurement at the same time.
Executes line/match calibration for test ports 3 and 4.
3-4 Fwd
Use this to perform forward-direction measurement
(S43)
only.
Executes line/match calibration for test ports 3 and 4.
3-4 Rvs
Use this to perform reverse-direction measurement
(S34)
only.
7. Press Done to finish TRL 4-port calibration. At this point, the
calibration coefficient is calculated and saved. The error correction
function is automatically turned on.
Measurement
263