Agilent Technologies E5071C Manual page 413

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
Setup Description
Activating Trace 2
Marker 1: ON
Search/Tracking: ON
Moving Marker 1 to the trace
maximum
Bandwidth search: ON
The insertion loss (loss) in this example is -1.627 dB and the 3-dB
bandwidth (BW) is 71.04 MHz.
3. Determine the isolation between Tx and Rx.
Setup Description
Activating Trace 3
Marker 1: ON
Search/Tracking: ON
Moving Marker 1 to the peak
near 1.92 GHz
The isolation (response value of marker 1) in this example is -6.612
dB.
4. Determine the return loss of Tx.
Setup Description
Activating Trace 4
Marker 1: ON
Search/Tracking: ON
Moving Marker 1 to the peak
in the pass band
The return loss (response value of Marker 1) in this example is 12.65
dB.
5. Determine the return loss of Rx.
412
Key Operation
Trace Next
Marker
Marker Search
ON.)
Max
Bandwidth (Turn it ON)
Key Operation
Trace Next
Marker
> Tracking (Turn it ON)
Marker Search
Peak - Search Left or Search Right (
press as many times as necessary)
Key Operation
Trace Next
Marker
> Tracking (Turn it ON)
Marker Search
Peak > Search Left or Search Right (
press as many times as necessary)
> Tracking (Turn it