5-5. Trace Status Area
The setup for each trace is displayed here.

Trace status display

Classification
Error correction
(In the option TDR,
this information
should be ignored.)
Turning on/off
traces
Performing data
math
38
Contents
Meaning
inside [ ]
Error correction: ON
RO
(OPEN (n) response
calibration)
Error correction: ON
RS
(SHORT (n) response
calibration)
Error correction: ON
RT
(THRU (n) response
calibration)
Error correction: ON
ER
(Enhanced response
calibration)
Error correction: ON (1-
F1
port calibration )
Error correction: ON (Full
F2
2-port calibration / 2-port
TRL calibration)
Error correction: ON (Full
F3
3-port calibration / 3-port
TRL calibration)
Error correction: ON (Full
F4
4-port calibration / 4-port
TRL calibration)
Nothing
Data trace: ON, Memory
trace: OFF
Data trace: OFF, Memory
M
trace: ON
Data trace: ON, Memory
D&M
trace: ON
Data trace: OFF, Memory
off
trace: OFF
D+M (D+M&M)
Execution of Data+Mem