Calibration type
1-Port Calibration
Scalar-Mixer
Calibration
Simply issuing one of the above commands completes some of the tasks
necessary for error correction, including measuring the calibration data,
calculating the calibration coefficients, and running the error correction
feature. To implement scalar-mixer calibration, you must measure the
power in advance. For 1-port calibration, you do not need to do this.
Once ECal has started, you cannot interrupt the operation.
Any command received after ECal has started will not be
executed until ECal is completed. In other words, even if you
issue a command with a Query, you cannot get the Query
response until ECal is finished.
1030
Command
:SENS{1-
36}:CORR:OFFS:COLL:ECAL:SOLT1
:SENS{1-
36}:CORR:OFFS:COLL:ECAL:SMIX2