Reading Actual Measurement Point/Value Interpolated between
Measurement Points
The point on the trace on which a marker can be placed differs depending on
how the discrete marker mode is set up.
Value
Description
Turning
A marker moves only between actual measurement points.
on
When a specific marker stimulus value is specified as a
discrete
numerical value, the marker is placed at the measurement
mode
point closest to the specified value. A marker that is placed
(Discrete
between interpolated points with the discrete mode off
automatically moves to the nearest measurement point
ON)
when the discrete mode is turned on.
Turning
The marker can move from one actual measurement point
off
to another. Because it is interpolated, it can also move in
discrete
the space between measurement points.
mode
(Discrete
OFF)
Turning Discrete Mode On or Off
Measurement
281