Agilent Technologies E5071C Manual page 91

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
2. If you change the on/off setting of the level coupling, all ports
are automatically changed to the same level value as that of
port 1.
3. When setting level for all ports (Port Couple ON), click Power input
the power level using the
When setting level for each port (Port Couple OFF), click Port Power
and select the softkey corresponding to each port (Port 1 Power to Port 4
Power).
The table below describes the relation between source power level and
step attenuation:
Source Power Range
-25 ~ +10 dBm
-35 ~ 0 dBm
-45 ~ -10 dBm
-55 ~ -20 dBm
-65 ~ -30 dBm
-75 ~ -40 dBm
-85 ~ -50 dBm
For Auto Power Range setting, the minimum value of step attenuation is
selected according to the available source power level. The table below
describes the step attenuation selection based on the setting source
power level:
Source Power Range
-25 ≤ power ≤ +10 dBm
-35 ≤ power < -25 dBm
-45 ≤ power < -35 dBm
-55 ≤ power < -45 dBm
-65 ≤ power < -55 dBm
-75 ≤ power < -65 dBm
-85 ≤ power < -75 dBm
90
block keys on the front panel.
ENTRY
Step Attenuation
0 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
Step Attenuation
0 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents