E5071C
54
Transform, Gate
not allowed
-211
Trigger ignored
U
No.
Message
300
Unable to
estimate
adapter length
-113
Undefined
header
774
characters exceeds 254 in a
character-string parameter.
This message appears when you turn
on the gating/conversion function of
the time domain feature, while
frequency offset is enabled,
measurement points are set to two,
or sweep type is set to other than
linear sweep
Turn off the frequency offset
(SENS:OFFS OFF command), set the
measurement points to three or more
and the sweep type to linear sweep,
and then turn on the
gating/conversion function.
This instrument receives and detects
a trigger command ("TRIG") or an
external trigger signal. But it is
ignored due to the timing condition
(This instrument is not in the wait-
for- trigger state, for example).
Change the setup so that a trigger
command or an external trigger
signal can be sent after the
instrument has entered the wait-for-
trigger state.
Description
Adapter length can not be estimated
at zero span. Change the frequency
span other than 0.
A command not defined in this
instrument, though not illegal in the
syntactic structure, has been
received. For example, when a
message ":DISP:WIND1:TABL:MEM
ON" is sent to a correct program
message ":DISP:WIND1:TRAC1:MEM
ON," the message sent is received as