Agilent Technologies E5071C Manual page 190

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

C1
C2
C3
L0
L1
L2
L3
Offset Delay
Offset Z0
Offset Loss
Arb.
Impedance
Min.
Frequency
Max.
Frequency
Media
Length Type
7.
Sets C1
Sets C2
Sets C3
Sets L0
Sets L1
Sets L2
Sets L3
Sets the offset delay
Sets the offset Z0. Specify 1 ohm when
Waveguide is set for the media type.
Sets the offset loss
Sets an arbitrary impedance
Sets a start frequency. Specify cutoff
frequency when Waveguide is set for the
media type.
Sets a stop frequency. Specify cutoff
frequency when Waveguide is set for the
media type.
Selects a media type, Coaxial or Waveguide.
When the standard type is Load, this softkey
can be selected from Fixed, Sliding or Offset.
Measurement
189