E5071C
Bit
Name
Position
0
Trace 15, 16 Limit Test
summary (questionable
limit channel {1-16} extra
status register summary)
1
Trace 1 Limit Test Fail
2
Trace 2 Limit Test Fail
3
Trace 3 Limit Test Fail
4
Trace 4 Limit Test Fail
5
Trace 5 Limit Test Fail
1128
Description
Set to "1" while one of the
enabled bits in the
questionable limit channel
{1-16} extra status event
register is set to "1."
Set to "0" when a
measurement cycle begins;
set to "1" when the
measurement cycle finishes
and returns "fail" as the limit
test result for trace 1.
Set to "0" when a
measurement cycle begins;
set to "1" when the
measurement cycle finishes
and returns "fail" as the limit
test result for trace 2.
Set to "0" when a
measurement cycle begins;
set to "1" when the
measurement cycle finishes
and returns "fail" as the limit
test result for trace 3.
Set to "0" when a
measurement cycle begins;
set to "1" when the
measurement cycle finishes
and returns "fail" as the limit
test result for trace 4.
Set to "0" when a
measurement cycle begins;
set to "1" when the
measurement cycle finishes
and returns "fail" as the limit
test result for trace 5.