calculated without a part of the thru measurement data. Therefore, the
following should be considered when measuring data for the simplified full
3-/4-port calibration.
The standard used for measurement must match its definition value.
Use a standard that provides good repeatability (stability).
Do not omit the length of the thru when defining the standard.
When using a user-created standard, verify the definition
value.
For the N connector, remember that it has two different types:
male and female.
Assure high reliability and repeatability for measurement.
Reduce the difference in external environment (such as
temperature difference) between the time when measuring
calibration data and when measuring actual data.
Set the power level of the stimulus signal sufficiently small so
that compression does not occur.
Narrow the IF bandwidth.
Increase the averaging factor.
Use a cable that has robust amplitude/phase characteristics
against bending.
Use high-precision connectors.
Simplified 3/4-port TRL calibration
This function is available with Firmware revision A.06.50 or
greater.
The simplified 3/4-port TRL calibration calculates the calibration
coefficient by skipping part of the thru/line measurement (or
line/match measurement) that is necessary for normal 3/4-port
TRL calibration.
You need to give consideration to the same conditions for the
simplified 3/4-port TRL calibration as those for the simplified full
3/4-port calibration listed above.
Turning ON/OFF Error Correction
To turn ON/OFF error correction, use the following command:
:SENS{1-160}:CORR:STAT
Programming
1013