E5071C
span
Peak not found
Segment table
changed
Sweep mode
changed to stepped
sweep
Target value not
778
frequency span (the stop frequency is about 4
times or more of the start frequency) when the
sweep type is set to the log sweep, this
message is displayed.
If this message is displayed, the sweep type is
automatically set to the linear sweep.
For example, if you attempt to change the log
type to the log sweep when the start frequency
is set to 1 MHz and the stop frequency to 2
MHz, this error occurs.
Set the stop frequency to a value of about 4
times or more the start frequency and then set
the low seep.
This warning message is displayed when, with
the tracking turned on, the peak specified by
the marker search has not been found by the
time the sweep is finished (with the tracking
executed).
This warning message is displayed when the
setting specified segment by segment in the
segment table is automatically changed by a
change in the other setting.
For example, this warning message is
displayed when, with the power specified
segment by segment in the segment table, the
power setting for a segment is adjusted by a
change in the power range setting.
This message appears when you set the sweep
type to power sweep, power slope function to
enabled (ON, correction coefficient is other
than zero), power calibration to ON, frequency
offset to enabled, spurious avoidance of
frequency offset to enabled, or external signal
source control to enabled, while the sweep
mode is set to swept mode. When this
message is displayed, the sweep mode is
automatically set to step mode (during swept
mode).