Create A Segment Sweep Table - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C

2. Create a Segment Sweep Table

Follow the steps below to make entries in the segment sweep table.
1. Display the segment table.
Setup Description
Presetting
Displaying the segment table
2. Display the IF bandwidth setting column in the segment table.
Setup Description
Moving the focus to the softkey menu
Display of IF bandwidth setting column: ON
When setup items (power level, delay time, sweep mode, and sweep
time in this case) are not displayed in the segment table, the setting
for the channel in use applies to all segments.
3. Enter the setup data in the segment table.
Setup Description
Moving the focus to the segment table
(select)
Segment 1
Start frequency: 440 MHz
Stop frequency: 915 MHz
Number of points: 47
IF bandwidth: 70 kHz
Segment 2
Start frequency: 915 MHz
Stop frequency: 980 MHz
396
Key Operation
> OK
Preset
> Edit Segment Table
Sweep Setup
Key Operation
Focus
List IFBW (turn it ON)
Key Operation
Focus
>
4
>
9
>
4
>
7
>
9
>
9
>
>
4
0
M/µ
>
>
1
5
M/µ
>
7
x1
>
0
k/m
>
>
1
5
M/µ
>
>
8
0
M/µ