Omissible Thru Type At 3-Port Trl Calibration; Omissible Measurements At 4-Port Trl Calibration - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

The types of thru/line measurement and line/match measurement you can
omit are determined by the ports you select. Some types of measurement
cannot be omitted. You cannot omit thru/line measurement only or
line/match measurement only, even for an omissible measurement path.

Omissible Thru type at 3-port TRL calibration

Ports used
Ports 1, 2, 3
Ports 1, 2, 4
Ports 1, 3, 4
Ports 2, 3, 4
For the simplified 3-port TRL calibration, (Optional) is displayed on the
softkey of the omissible thru/line measurement. For example, when the
omissible thru/line measurement is 2-3, 2-3 (Optional) is displayed. The
display after the execution of the omissible thru/line measurement is 2-3
Thru/Line.
For the softkey of an omissible line/match measurement, (Optional) is
displayed as in the case of the omissible thru/line measurement. For
example, when the omissible line/match measurement is 2-3, 2-3 (Optional)
is displayed. The display after the execution of the omissible line/match
measurement is 2-3 Line/Match.
The softkeys (Fwd and Rvs) for forward measurement and reverse
measurement for the omissible line/match measurement do not change.

Omissible measurements at 4-port TRL calibration

Required measurements
1-2, 1-3, 3-4
For the simplified 4-port TRL calibration, (Recommended) is displayed on
the softkey of the omissible thru/line measurement. For example, when
the omissible thru/line measurement is 2-3, 2-3 (Recommended) is
displayed. The display after the execution of the omissible thru/line
measurement is 2-3 Thru/Line, and the display for the remaining omissible
thru/line measurement is 2-3 (Optional).
For the softkey of an omissible line/match measurement, (Recommended)
is displayed as in the case of the omissible thru/line measurement. For
example, when the omissible line/match measurement is 2-3, 2-3
Omissible measurement
2-3
2-4
1-4
2-4
Omissible measurements
1-4, 2-3, 2-4
Measurement
221

Hide quick links:

Advertisement