Agilent Technologies E5071C Manual page 377

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
You can set the auto port extension function channel by
channel. Setting the auto port extension function for one
particular channel does not affect other channels.
When the sweep type is power sweep or the frequency offset
function is ON, the auto port extension is not available.
Auto Port Extension does not supports waveguide port
extension.
1. Selecting a port(s)
Select the port(s) for which you want to use the auto port extension
function.
1. Press
Channel Next/Channel Prev
which you want to set auto port extension.
2. Press
Cal
3. Click Port Extensions > Auto Port Extension > Select Ports to select
the port(s) for which you want to use the auto port extension
function.
2. Setting frequencies used for calculation
Set the frequency points with which you want to calculate a loss value.
1. Press
Channel Next/Channel Prev
auto port extension.
2. Press
Cal
3. Click Port Extensions > Auto Port Extension > Select Ports > Method
to set the frequencies used for calculation.
Softkey
Current
Span
Active
Marker
User Span
376
key.
key.
Function
Executed using the frequency range set currently.
Executed using the frequency at the
In this case, the result is applied to Loss1. Loss2 is
ignored.
Executed using a start value and a stop value you
set.
keys to activate the channel for
keys to activate the channel to set
active
marker.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents