Connecting The Standard For 1-Port Calibration - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
7. Connect a SHORT calibration standard to the selected test port
(connector to which the DUT is to be connected).
8. Click Short to start the calibration measurement.
9. Connect a LOAD calibration standard to the selected test port
(connector to which the DUT is to be connected).
10.
Click Load to start the calibration measurement.
11.
Click Done to terminate the 1-port calibration process. Upon
pressing this key, calibration coefficients will be calculated and
saved. The error correction function will also be automatically
enabled.

Connecting the standard for 1-port calibration

138