3100 Segm=2 ! Number of Segment Ch.1 : 2
3110 Star1(1)=3.E+6 ! Start Frequency Ch.1 Segm.1: 3.0 MHz
3120 Star1(2)=5.0E+7 ! Segm.2: 50.0 MHz
3130 Stop1(1)=1.0E+7 ! Stop Frequency Ch.1 Segm.1: 10.0 MHz
3140 Stop1(2)=8.E+9 ! Segm.2: 8.0 GHz
3150 Nop1(1)=2 ! Number Ch.1 Segm.1: 2
3160 Nop1(2)=170 ! of Points Segm.2: 170
3170 If_bw1(1)=7.0E+4 ! IF Bandwidth Ch.1 Segm.1: 70 kHz
3180 If_bw1(2)=7.0E+4 ! Segm.2: 70 kHz
3190 Pow1(1)=0 ! Power Ch.1 Segm.1: 0 dBm
3200 Pow1(2)=0 ! Segm.2: 0 dBm
3210 !
3220 OUTPUT @Agte507x;":SYST:PRES"
3230 !
3240 WAIT 5
3250 !
3260 ! Channel 1
3270 !
3280 OUTPUT @Agte507x;":SENS1:SWE:TYPE SEGM"
3290 OUTPUT @Agte507x;":SENS1:SEGM:DATA 5,0,1,1,0,0,";Segm;",";
3300 FOR I=1 TO Segm-1
3310 OUTPUT @Agte507x;Star1(I);",";Stop1(I);",";Nop1(I);",";If_bw1(I);",";Pow1(I);",";
3320 NEXT I
3330 OUTPUT
@Agte507x;Star1(Segm);",";Stop1(Segm);",";Nop1(Segm);",";If_bw1(Segm);",";Pow1(Segm)
3340 !
3350 OUTPUT @Agte507x;":CALC1:PAR:COUN ";Num_of_tr1
3360 FOR I=1 TO Num_of_tr1
3370 OUTPUT @Agte507x;":CALC1:PAR"&VAL$(I)&":SEL"
3380 NEXT I
3390 SUBEND
3400!
3410 Select_cal_kit: SUB Select_cal_kit(@Agte507x,Ch$)
3420 !=============================================
3430 ! Calibration Kit Selection Function
3440 !=============================================
3450 !
1206