106
Save failed
33
Selected
parameter not
valid for
confidence
check
57
SnP request not
valid for
selected
measurement
50
Specified
channel hidden
23
Specified error
term does not
exist
21
Specified ports
overlapped
This error occurs when writing an
instrument status file (State01.sta,
etc.) (MMEM:STOR:STAT command)
fails.
This error occurs when the mix mode
S parameter has been selected for the
S parameter you want to check, while
using the confidence check function
for calibration coefficient.
This error occurs when you try to save
data to a Touchstone file but no
measurement has been executed.
This error also occurs when you try to
save a Touchstone file with power
sweep measurement specified or with
the frequency offset function set to
ON.
This error occurs when an attempt is
made to activate a channel not on
display using the DISP:WIND:ACT
command. This error is not generated
by front key operations.
The error occurs when the calibration
coefficient type of the parameter
specified in the read/write command
of the calibration coefficient
(SENS:CORR:COEF) is invalid for the
calibration type selected by the
calibration type selection command
(SENS:CORR:COEF:METH:xxxx.) This
error is not generated by front key
operations.
This error occurs when a port number
is duplicated in a command requiring
two or more port numbers as
parameters. Such commands are, for
example,
CALC:FSIM:BAL:TOP:SSB:PPOR
1,2,3,3. Specify port setup correctly to
Product Information
771
Need help?
Do you have a question about the E5071C and is the answer not in the manual?
Questions and answers