Reflection tracking error (Er)

A reflection tracking error is caused by the difference in frequency
response between the test receiver and the reference receiver of a
stimulus port in reflection measurements. This error can be defined as a
constant value for each stimulus port because the combination of the test
receiver and the reference receiver of a stimulus port is always the same.
The number of reflection tracking errors for the E5071C is simply the
number of stimulus ports you use.
Er1
Reflection tracking error of port 1
Er2
Reflection tracking error of port 2
Er3
Reflection tracking error of port 3
Er4
Reflection tracking error of port 4

Transmission tracking error (Et)

A transmission tracking error is caused by the difference in frequency
response between the test receiver of a response port and the reference
receiver of a stimulus port in transmission measurements. When a certain
port is a stimulus port, a transmission tracking error is defined for each of
the other ports. Therefore, the number of transmission tracking errors for
the E5071C is the total number of combinations of stimulus ports and
response ports.
Et21, Et31, and Et41
Et12, Et32, and Et42
Et13, Et23, and Et43
Et14, Et24, and Et34
Transmission tracking error when port 1 is a stimulus
port
Transmission tracking error when port 2 is a stimulus
port
Transmission tracking error when port 3 is a stimulus
port
Transmission tracking error when port 4 is a stimulus
port