:SENS{1-160}:CORR:COLL:ECAL:ISOL
This command takes no action and only exists to maintain
backward compatibility.
To select the ECal characteristic for a user-characterized ECal, use the
following command:
:SENS{1-160}:CORR:COLL:ECAL:UCH
ECal Auto-detect Function
The ECal module can automatically detect which port of the ECal module is
connected to the E5071C test port. Turn off the auto-detect function to
specify a port manually.
To turn OFF the auto-detect function, use the following command.
:SENS:CORR:COLL:ECAL:ORI
To turn OFF the auto-detect function and set a port manually, use the
following command.
:SENS:CORR:COLL:ECAL:PATH
Checking the Applied Calibration Type
When you turn on error correction, you can check the calibration type
actually applied to each trace. To check the calibration type, use the
following command:
:SENS{1-160}:CORR:TYPE{1-160}?
The above command reads out the same parameter (SOLT3) for both the
full 3-port and simplified full 3-port calibrations, and thus they cannot be
discriminated. For the same reason, the following calibrations cannot be
discriminated: full 4-port and simplified full 4-port, 3-port TRL and
simplified 3-port TRL, and 4-port TRL and simplified 4-port TRL.
Defining Calibration kits
Selecting a Calibration Kit
To select a calibration kit, use the following command:
:SENS{1-160}:CORR:COLL:CKIT
Setting the Calibration Kit Name
To set the name of a calibration kit, use the following command:
:SENS{1-160}:CORR:COLL:CKIT:LAB
Standard Definitions
Selecting a Standard Type
Programming
1015