Clearing Calibration Data and Calibration Coefficients

Clearing Calibration Data
When the frequency offset function has been disabled
You can use the following command to clear the measurement values of
calibration data executed with :SENS{1-160}:CORR:COLL:OPEN
command, etc.
:SENS{1-160}:CORR:COLL:CLE
When the frequency offset function has been enabled
You can use the following command to clear the measurement values of
calibration data executed with :SENS{1-160}:CORR:OFFS:COLL:OPEN
command, etc.
:SENS{1-160}:CORR:OFFS:COLL:CLE
These clear functions make the temporarily settings during the calibration,
such as trace number and measurement parameters, recover to the
original state.
Clearing Calibration Coefficients
When the frequency offset function has been disabled
You can use the following command to clear the calibration coefficients
used.
:SENS{1-160}:CORR:CLE
When the frequency offset function has been enabled
You can use the following command to clear the calibration coefficients
used.
:SENS{1-160}:CORR:OFFS:CLE
This command does not clear calibration coefficients related to normal
calibration.
Programming
1021