E5071C
To select whether to save the setting of all channels/traces or that of the
displayed channels/traces, use the following command:
:MMEM:STOR:SALL
Saving and recalling entire instrument status
To save the entire instrument status into a file, use the following
command:
:MMEM:STOR
Recalling a file saved with the above command can reproduce the status
when it was saved. To recall the settings from a file, use the following
command:
:MMEM:LOAD
Auto recall
The file saved with the name autorec.sta or A:autorec.sta will be
automatically recalled when the E5071C is powered ON.
Saving the state for each channel into a register
For the active channel, when you want to save the instrument state
specific to that channel into only one of the registers A to D, use the
following command:
:MMEM:STOR:CHAN
Recalling an instrument state saved in a register can reproduce it as the
state of the active channel. To recall a register, use the following
command:
:MMEM:LOAD:CHAN
It is possible to recall a file from a different channel where it
•
was saved.
The contents in the registers are lost when you turn OFF the power. You
can delete (clear) the contents of all registers using the following
command.
:MMEM:STOR:CHAN:CLE
Saving Measurement Data
Measurement data (in a formatted data array) can be saved to a file in
CSV (Comma Separated Value) format.
To save measurement data in a file, use the following command:
:MMEM:STOR:FDAT
1086