E5071C
To select whether to save the setting of all channels/traces or that of the
displayed channels/traces, use the following command:
:MMEM:STOR:SALL
Saving and recalling entire instrument status
To save the entire instrument status into a file, use the following
command:
:MMEM:STOR
Recalling a file saved with the above command can reproduce the status
when it was saved. To recall the settings from a file, use the following
command:
:MMEM:LOAD
Auto recall
The file saved with the name autorec.sta or A:autorec.sta will be
automatically recalled when the E5071C is powered ON.
Saving the state for each channel into a register
For the active channel, when you want to save the instrument state
specific to that channel into only one of the registers A to D, use the
following command:
:MMEM:STOR:CHAN
Recalling an instrument state saved in a register can reproduce it as the
state of the active channel. To recall a register, use the following
command:
:MMEM:LOAD:CHAN
It is possible to recall a file from a different channel where it
was saved.
The contents in the registers are lost when you turn OFF the power. You
can delete (clear) the contents of all registers using the following
command.
:MMEM:STOR:CHAN:CLE
Saving Measurement Data
Measurement data (in a formatted data array) can be saved to a file in
CSV (Comma Separated Value) format.
To save measurement data in a file, use the following command:
:MMEM:STOR:FDAT
1086