E5071C
5. Click Reference Plane to select the
standard
Softkey
Thru
Reflect
6.
Setting a media type for the calibration kit
You can set a media type for the standard you use.
1. Press
Cal
2. Click Cal Kit, then select a calibration kit.
3. Click Modify Cal Kit > Modify Cal Kit.
4. Click Define STDs, and select a standard.
5. Click Media, and select a
Softkey
Coaxial
Waveguide
6.
7. If you select waveguide as the media type, set the system
impedance and the characteristic impedance to 1ohm .
8. Depending on the media type you use, the calculation method
of the electrical delay, which is required to correct the phase
delay, differs.
194
test port impedance differ.
used for calculating the reference plane.
Function
Uses the length of the thru/line standard to calculate the
calibration plane.
Uses the reflection coefficient of the reflection standard
to calculate the calibration plane.
key.
media
Function
Selects coaxial as the media type.
Selects waveguide as the media type.
measurement data of the
type.