Calibration And Measurement Parameters; Calibration When Using The External Test Set - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

2. Mode 2
Setting procedure
Follow these steps to set the external test set mode.
1. Press
System
2. Select your desired mode.
3. When the setting of the external test set mode is changed, the
calibration coefficients acquired using the calibration kit or ECal
are cleared, and the error correction function is turned off.

Calibration and Measurement Parameters

When the external test set mode (Mode 1 or Mode 2) is used with the
E5071C, only limited types of calibrations and measurement parameters
are available.

Calibration when using the external test set

When the external test set mode (Mode 1 or Mode 2) is used with the
E5071C, the following calibrations are available. When other calibration
methods are used, correct calibration is not performed.
Open response calibration
Short response calibration
Thru response calibration
Enhanced response calibration
1-port calibration
When ECal is used with Mode 1 or Mode 2 selected, the
connection of test port 4 cannot be detected automatically.
When you perform calibration using the 4-port ECal, set the
connection of port 4 as follows depending on the connection of
test port 1/2/3.
> Service Menu > External Test Set.
Setting Control Functions
739