Math Operation Functions On Traces; Device Test Functions; Analytical Functions - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
channel also supports a fixed marker that can be established at a fixed
position.
In contrast, the E5071C enables you to have all 36 channels displayed by
opening up 36 separate windows on the screen. Sixteen traces can be
displayed for each channel, and up to nine markers can be displayed for
each trace. In addition to the markers displayed, you can also designate
one marker as the reference marker. The E5071C, however, does not
support fixed markers.
The 8753ES supports the Maximum, Minimum, Target value, and
Bandwidth marker functions. The E5071C supports all these in addition to
a Peak Search function. Using this function, you can determine whether or
not to search for a positive or negative peak. In addition, the 8753ES has
an additional function to search for the maximum or minimum bandwidth.
While the 8753ES allows a target value or search tracking to be established
only on the active marker, the E5071C allows a target value or search
tracking to be established on all markers.

Math Operation Functions on Traces

On the 8753ES, each channel is provided with a memory trace. For this
reason, math operations between the data trace and memory trace are
supported: Data / Memory and Data - Memory On the E5071C, however, Data
* Memory and Data + Memory are supported along with the division and
subtraction operations described above.
The trace displays supported on the 8753ES are: "Data trace only",
"Memory trace only", "Both memory trace and data trace" and "Data math
only". The E5071C supports these functions in addition to "Both data math
and memory trace"

Device Test Functions

The 8753ES supports limit test, ripple test and bandwidth test. The E5071C
also supports these tests.
For the limit test on the 8753ES, a limit test table is provided for each
channel with up to 18 segments are allowed in each table. In contrast, the
E5071C allows a limit test table to be defined for each trace and up to 100
segments to be defined per table.
The 8753ES supports the pass/fail result for each channel, segment or
measurement point along with the minimum/maximum measurement
value for each segment as retrieved items in test results. On the other
hand, the E5071C supports pass/fail results, measurement value, and
upper/lower limit values for each measurement point as well as pass/fail
results of the active trace for each channel.

Analytical Functions

908