Characterizing Calibration Mixer Using Ecal - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

where the up/down conversion method is used, the power of the
LO signal is distributed to the calibration mixer and the
measured mixer through the power splitter. During a
characteristics evaluation of the calibration mixer, the LO power
level used by the drive of the calibration mixer must be equal to
the LO power level with the measured mixer connected. This is
because the mixer's conversion loss and reflection coefficient
are significantly affected by the power level of the LO signal.
1. Select CalKit (3 in menu).
2. Select the type number of the calibration kit from CalKit menu (4 in
menu).
3. Click the Open button (5 in menu) to start measurement in OPEN.
4. Click the Short button (6 in menu) to start measurement in SHORT.
5. Click the Load button (7 in menu) to start measurement in LOAD.
7. Measuring Calibration Mixer with IF Filter (when using ECal module)

Characterizing calibration mixer using ECal

1. Select ECal (1 in menu).
2. Select the port used for the ECal module (2 in menu).
3. Click the Measure button (3 in menu) to start measurement.
8. Saving Characteristic Data and Closing VBA Macro
Measurement
251