Setting Trace Display - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
From Firmware revision A.08.10, channel display configuration
of 8×9, 6×12 and 8×12 is added.

Setting Trace Display

Setting the number of traces
Depending on the measurement parameters of the traces displayed for
each channel, the sweep necessary for each channel is executed. For more
information, refer to Sweep Order in Each Channel.
You specify the trace display by setting the number of traces (upper limit
of displayed trace numbers). For example, if you set the number of traces
to 3, traces 1 through 3 are displayed.
The procedure for setting the number of traces is as follows:
1. Press
Channel Next
you want to set the number of traces.
2. Press
Display
3. Press the desired softkey to set the number of traces.
Setting trace layout (graph layout)
Traces are laid out and displayed in the order of the trace number from
graph 1 according to the graph layout in the channel window.
You can select the graph layout from the windows layout.
If the number of traces is less than the number of graphs, nothing is
displayed in the remaining area. If the number of traces you set exceeds
the number of graphs, excess traces are superimposed from the first
graph. For example, if you select
the number of traces to 5, graph 1 (Gr1 in
in
Graph
layout) display traces 1 and 4 and traces 2 and 5, respectively, by
superimposing, and graph 3 (Gr3 in
shown in the figure below.
76
or
Channel Prev
> Number of Traces.
Graph
to select the channel for which
as the graph layout and set
Graph
layout) and graph 2 (Gr2
layout) displays only trace 3 as

Hide quick links: