Limit Line Offset - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Softkey
Limit
Test
Limit
Line
Edit Limit
Line

Limit line offset

By adding a certain offset to the limit value, you can adjust the limit line so
that it conforms to the device output.
1. Press
Channel Next/Channel Prev
to select the trace on which the limit test function will be used.
2. Press
Analysis
3. Click Limit Test to display the softkeys for the limit test.
4. Click Limit Line Offsets to display the limit line offset function menu.
The following functions correspond to each softkey.
Softkey
Stimulus
Offset
Amplitude
Offset
Marker ->
Amp. Ofs.
Stimulus offset
Function
Sets the limit test ON/OFF.
Sets the limit line display ON/OFF.
Opens the limit table for editing the limit line. See
Defining_the_limit_line.
key to display the Analysis menu.
Function
Adds a certain offset to the stimulus value of the
entire segment in the limit table. (Stimulus offset)
Adds a certain offset to the response value of the
entire segment in the limit table. (Amplitude
offset)
Adds the amplitude offset by the same amount as
the retrieved value of the
confirm the current value set for the amplitude
offset by pressing Amplitude Offset. (Marker
amplitude offset)
keys and
Trace Next/Trace prev
active marker
Measurement
keys
. You can
313