Although the 8753ES does not support the fixture simulator function, the
E5071C does. The fixture simulator supported by the E5071C include the
balanced-unbalanced transformation function for analyzing balanced
devices, and the matching circuit function.
Save/Recall
For storing data, the 8753ES is provided with an internal register, internal
disk drive (floppy disk), and external disk drive (connected through the
GPIB). In contrast, the E5071C provides an internal hard disk drive and
allows external storage devices such as USB memory. While the 8753ES
can save or recall the device setup, screen color settings, and test
sequences, the E5071C is able to save or recall the instrument setup,
segment sweep table, and limit line table as well as VBA projects. The
8753ES saves display data in JPEG format while the E5071C supports the
Windows Bitmap (BMP) and Portable Network Graphics (PNG) format.

Test Sequence Program

Although the 8753ES supports the test sequence program, the E5071C
provides an environment for developing VBA programs for automatic
measurement.

Outputting to a Printer/Plotter

The 8753ES enables you to establish the print area covering an entire
sheet or just a 1/4 sheet, and to define traces, the reference line, and
colors of warning messages. In contrast, the E5071C supports only full-
sheet output, and an on/off setting for highlighting the entire screen in
connection with color setup.
The 8753ES supports parallel ports, serial ports, and GPIB as printer ports,
although the E5071C supports only USB-connected printers.

GPIB Interface

While the 8753ES uses pass control to pass the controller information to an
external PC or instrument, the E5071C does not support this function.

LAN Interface

Although the 8753ES does not support a LAN interface, support for LANs is
standard on the E5071C.
Other Functions
The 8753ES is provided with Take4 mode, mixer measurement function,
and harmonics measurement function (Option 002). The E5071C provides
the mixer measurement function, and harmonics measurement function
with option 008.