E5071C
-
File name
257
error
-
File name
256
not found
H
No.
Message
-114
Header suffix
out of range
I
No.
Message
-224
Illegal
parameter
value
601
Invalid bit
pattern
-161
Invalid block
data
28
Invalid
calibration
method
782
A file name error. This message appears when an
error exists in the file name and hence a
command is not executed correctly. This error
occurs when you try to copy to an unsuitable file
name, for example.
The file name specified is not found and hence the
command is not executed correctly. This error
occurs when you try to read a file that does not
exist in a disk or a disk is not correctly inserted
into the drive to read or write a file, for example.
Description
The unit of the header is outside the range.
The header is invalid in the unit for numeric
parameters following a SCPI command.
Description
The parameter value is not suitable. This
error occurs when the CALC:TRAC:PAR
command is used to specify an S-parameter
that does not exist in the model (S44 in the
case of a 2-port model), for example.
An invalid bit pattern is entered or loaded.
Block data has been expected, but the block
data that appears is invalid for some reason
(see 7.7.6.2, IEEE488.2). The END message
is received before the length of block data
has been filled, for example.
This error occurs when execution of load
measurement
(:SENS:CORR:COLL:DLC:LOAD), open
measurement
(:SENS:CORR:COLL:DLC:OPEN), thru
measurement as part of DLC sequence
(:SENS:CORR:COLL:DLC:THRU) or