Connecting 4-Port Ecal Module (For Full 4-Port Calibration) - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
9.
10.
If you must select a port, the softkey for making this selection is
displayed. Select a port and start calibration. If you do not have to
select a port, this step is skipped.
11.
The E5071C detects the test ports connected to the ECal and
then measurement starts. If the selected test ports to be calibrated
is not connected to the ECal module, an error occurs.

Connecting 4-port ECal module (for full 4-port calibration)

You can connect the ports of the ECal and the test ports of the
E5071C arbitrarily. Connected ports can be manually specified
although they are automatically detected before data
measurement. For more information, Turning off ECal auto-
detect function.
156