result unless the frequency is the same for both the receiver
port and the source port.
When using power calibration in combination with receiver
calibration, you must implement it so that it sufficiently covers
the frequencies of both ports or implement it twice, before and
after the receiver calibration.
Turning ON/OFF Receiver Error Correction
To turn on or off the receiver error correction, use the following command:
:SENS{1-36}:CORR:REC{1-4}
Once the measurement of receiver calibration data is completed
with the
receiver error correction is automatically turned on.
:SENS{1-36}:CORR:REC{1-4}:COLL:ACQ
Programming
command,
1027