result unless the frequency is the same for both the receiver
port and the source port.
When using power calibration in combination with receiver
•
calibration, you must implement it so that it sufficiently covers
the frequencies of both ports or implement it twice, before and
after the receiver calibration.
Turning ON/OFF Receiver Error Correction
To turn on or off the receiver error correction, use the following command:
:SENS{1-36}:CORR:REC{1-4}
Once the measurement of receiver calibration data is completed
•
with the
receiver error correction is automatically turned on.
:SENS{1-36}:CORR:REC{1-4}:COLL:ACQ
Programming
command,
1027