1. Press
Channel Next/Channel Prev
you want to perform the calibration.
2. Press
key.
Cal
3. Click Calibrate.
4. Click Enhanced Response.
5. Click Ports to select the test ports on which an enhanced response
calibration will be performed.
6. Connect an OPEN calibration standard to the output port.
7. Click Open to start the calibration measurement.
If you select the calibration kit which has different
o
calibration definitions for each gender, (m) and (f) in the
name (label) of the standard displayed in the softkey
indicate male (m) and female (f) for the analyzer's
connector, respectively.
8. Disconnect the OPEN calibration standard and replace it with a
SHORT calibration standard.
9. Click Short to start the calibration measurement.
10.
Disconnect the SHORT calibration standard and replace it with a
LOAD standard.
11.
Click Load to start the calibration measurement.
12.
Make a THRU connection between the two ports.
13.
Click Thru to start the calibration measurement.
14.
If an isolation calibration must be performed using a LOAD
standard, follow the procedure below:
13.
a. Connect a LOAD standard to the two test ports.
keys to select the channel for which
Measurement