Agilent Technologies E5071C Manual page 603

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
The System Cannot be Operated
Manually (Front Panel Keys,
Keyboard, Touch Screen and
Mouse)
The keyboard or mouse becomes
inoperable.
The front panel key or keyboard
becomes inoperable.
The touch screen becomes
inoperable.
The mouse becomes inoperable.
All of the front panel keys,
keyboard, and mouse become
inoperable.
The keyboard and mouse have
been connected after power-on.
830
signal output is turned on.
Confirm that the keyboard or
mouse is connected
correctly. When it is
connected correctly, turn off
the power once, and restart
the system.
Using the mouse, turn
System
> Key Lock > Front
Panel & Keyboard Lock OFF.
Using the front panel keys,
press
> Key Lock >
System
Touch Screen & Mouse Lock
OFF to turn OFF the lock.
Execute the calibration of the
touch screen.
After the System Recovery,
touch screen may be
inoperable. See the
Reconfiguring Touch Panel
Setting in the installation
guide.
Using the front panel keys,
press
> Key Lock >
System
Touch Screen & Mouse Lock
to turn OFF the lock.
Confirm that the keyboard or
mouse is connected
correctly. When it is
connected correctly, turn off
the power once, and restart
the system.
Turn off the power once, and
restart the system. When
taking all these measures
does not recover operability,
there is a possibility of a
failure.