Measuring the Deviation from a Linear Phase

Overview
Procedure
Other Measurement Examples
Overview
This section illustrates how to determine the deviation from a linear phase
in the pass band of a 1.09-GHz bandpass filter.
Procedure
Here, the DUT is evaluated by following the steps.
Step
1. Connect the DUT
2. Define the
Measurement Conditions
3. Execute the
Calibration
4. Connect the DUT and
Execute the Auto Scale
5. Specify the Electrical
Delay
6. Measure the Deviation
from a Linear Phase
1. Connect the DUT
Connect the DUT as shown below.
418
Description
The DUT is connected.
The measurement conditions are defined.
The calibration is executed.
The DUT is connected again to execute the
auto scale function.
The electrical delay is specified.
The statistics data function (peak-to-peak)
is used to determine the deviation from a
linear phase.