Agilent Technologies E5071C Manual page 219

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

E5071C
5. Click Transmission.
6. Make a thru connection between selected test ports x and y
(between the connectors to which the DUT is connected).
7. Click Port x-y Thru (x and y are the thru-connected port numbers) to
start the re-measurement of the calibration standard.
8. Click Return.
9. Click Overwrite to finish the re-calibration for the full 2-port
calibration. At this point, the calibration coefficients are re-calculated
and saved.
.
218