E5071C
6. Make the full 2-port calibration measurement DONE. The calibration
factor is calculated based on the calibration data acquired, and the
error correction is turned ON.
> Calibrate > 2-Port Cal > Done
Cal
7. Select the type in which the data is to be saved before saving the
calibration factor (calculated based on the calibration data).
> Save Type > State & Cal
Save/Recall
8. Store the calibration file to the disk of the E5071C. The symbol "X"
appearing in the operations below represent the assigned numbers to
be used when the file is saved.
> Save State > State 0X
Save/Recall
STEP 3. Connecting the Device Under Test (DUT)
1. Connect to the DUT to the E5071C. (See the below figure)
Connecting the DUT
62