Advertisement

4-port TRL Calibration

Overview
Procedure
Other topics about TRL Calibration
Overview
The 4-port TRL calibration function lets you measure calibration data by
connecting thru, reflection (open or short), line, or match calibration
standards to (between) 4 test ports. This calibration provides the most
accurate measurement for non-coaxial parts, using 48 error terms model
in total for calibration in the same way as full 4-port calibration.
Before executing TRL calibration, you need to modify the calibration kit
definition you use (or create a new one). For more information, see
Changing the Calibration Kit Definition.
Procedure
1. Press
Channel Next/Channel Prev
you want to execute calibration.
2. Press
Cal
3. Click Calibrate > 4-Port TRL Cal > Select Ports.
4. Click Thru/Line to start the measurement of the calibration standard.
5. Click Reflect to start the measurement of the calibration standard.
Softkey
Port 1 Reflect
Port 2 Reflect
Port 3 Reflect
Port 4 Reflect
6. Click Line/Match to start the measurement of the calibration
standard.
Softkey
1-2
Line/Match
1-2 Fwd
key.
Function
Executes reflection calibration for test port 1.
Executes reflection calibration for test port 2.
Executes reflection calibration for test port 3.
Executes reflection calibration for test port 4.
Function
Executes line/match calibration for test ports 1 and
2.Use this to perform forward-direction measurement
and reverse-direction measurement at the same time.
Executes line/match calibration for test ports 1 and
keys to select the channel for which
Measurement
261