Compatibility When The Maximum Number Of Channels/Traces Is Different; Compatibility When The Firmware Revision Is Different; Compatibility When The System Spec. Version Is Different; Saving Data - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Option:
number
of port
Saved
2 Ports
File
4 Ports
by
Y: Recall is possible
N: Recall is impossible
Y/ c: Only when the stored data is "state only (State Only)," recall is
possible.

Compatibility when the maximum number of channels/traces is different

You cannot recall files saved by specifying all channels/traces (All) as
the save target.
If the number of channels/traces at save does not exceed that at
recall, you can recall files saved by specifying the displayed
channel/trace ( Disp Only) as the save target.

Compatibility when the Firmware revision is different

Files that are saved with later firmware revisions cannot be recalled
by prior firmware revisions.
Recalling the state file which is saved by prior firmware revision may
cause a sweep time change and C? is displayed. In this case, re-
calibration is recommended.
Compatibility when the system spec. version (available with ":SERV:SREV?") is different
When Files that are saved with different system spec. versions
include calibration data, only states and trace data are recalled.
If you recall an incompatible file, an error occurs and the device
recovers to the presetting.

Saving Data

Selecting Content to be Saved
This setting takes effect both when saving the entire instrument
state into a file and when saving the instrument state for each
channel into memory.
1. Press
Save/Recall
2. Click Save Type.
3. Click the softkey corresponding to the content of the instrument
state you want to save.
Selecting Save Target Channel/Trace
1. Press
Save/Recall
2 Ports
4
Ports
Y
Y/ c
N
key.
key.
Y
Measurement
345