Connecting Standards In Full 4-Port Calibration - Agilent Technologies E5071C Manual

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

12.
Connect a LOAD standard to each of the four test ports
(connectors to which the DUT is to be connected).
13.
Click Port 1-2 Isol to start the calibration measurement.
14.
Perform the each calibrations by clicking Port 1-3 Isol, Port 1-4
Isol, Port 2-3 Isol and Port 2-4 Isol.
15.
Click Return.
16.
Click Done to terminate the full 4-port calibration process. Upon
pressing this key, calibration coefficients will be calculated and
saved. The error correction function will also be automatically
enabled.

Connecting standards in full 4-port calibration

Measurement
147