E5071C

5. Setting the Time Domain Function

Set the conversion function to display the response in the time domain. If
you enable this setting, the response in time domain is displayed as shown
figure below. A peak indicating a small mismatch appears at the location of
the connector.
Setting Description
Data format: real
Setting the transformation type to
low-pass impulse
Setting the window type to maximum.
Setting the display range: from 0 s to
20 ns
Enabling the transformation function
Executing auto scale
436
Key Operation
> Real
Format
> Transform > Type >
Analysis
Lowpass Imp
Window > Maximum
Start >
>
0
x1
Stop >
>
>
2
0
G/n
Transform (set to ON)
> Auto Scale
Scale