Agilent Technologies E5071C Manual page 130

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

2. Press
key.
Cal
3. Click Calibrate.
4. Select
OPEN or SHORT response
Softkey
Response
(Open)
Response
(Short)
5.
6. Click Select Port.
7. Select the test port upon which you will perform OPEN/SHORT
response calibration.
8. Connect a calibration standard (OPEN or SHORT) to the selected test
port (connector to which the DUT is to be connected).
9. Click Open or Short to start the calibration measurement.
10.
If an isolation calibration must be performed using a LOAD
standard, follow the procedure below.
11.
Connect a LOAD standard to the selected test port (connector to
which the DUT is to be connected).
12.
Click Load (Optional) to start the measurement on the LOAD
standard.
13.
Click Done to terminate the response calibration (and the LOAD
isolation calibration) process. Upon pressing this key, calibration
coefficients will be calculated and saved. The error correction
function will also be automatically enabled.
Connecting standards in OPEN/SHORT Response calibration
Function
Displays softkeys for performing an OPEN response
calibration (response calibration with an OPEN
standard)
Displays softkeys for performing a SHORT response
calibration (response calibration with a SHORT
standard)
calibration.
Measurement
129