Saving and loading calibration coefficients
You can save calibration coefficients to a file along with other instrument
settings and then later load them from the file.
By default, the system does not save calibration coefficients when it saves
instrument settings. Therefore, to save calibration coefficients, you must
explicitly configure the system to save them by issuing the following
command:
:MMEM:STOR:STYP
To save calibration coefficients to a file, use the following command:
:MMEM:STOR
To load calibration coefficients from a file, use the following command:
:MMEM:LOAD
For more information on how to save and load calibration coefficients, refer
to Saving and recalling instrument status
Reading/Writing Calibration Coefficient Alone
The calibration coefficient alone can be read from and written to the
E5071C by using the following command:
:SENS{1-160}:CORR:COEF
To write a positive calibration coefficient, use one of the following
commands to declare the calibration type:
:SENS{1-160}:CORR:COEF:METH:ERES
:SENS{1-160}:CORR:COEF:METH:OPEN
:SENS{1-160}:CORR:COEF:METH:SHOR
:SENS{1-160}:CORR:COEF:METH:SOLT1
:SENS{1-160}:CORR:COEF:METH:SOLT2
:SENS{1-160}:CORR:COEF:METH:SOLT3
:SENS{1-160}:CORR:COEF:METH:SOLT4
:SENS{1-160}:CORR:COEF:METH:THRU
To validate the written calibration coefficient, use the following command:
:SENS{1-160}:CORR:COEF:SAVE
About Calibration Types and Coefficients
The following table shows the required calibration coefficients for each
calibration type.
Calibration Type
Calibration Coefficient
Programming
1019